5-7
4.
Press button for
Center Pause
: adjust to highlight 0, 2, 5, 10, 20 or 30 min.
Center pause is the amount of time the patient is held in a level position after rotating to the right or left.
5.
Press button for
Rotate Hour Meter
: zeroes the meter which measures the time spent in rotation mode.
6.
Press button for
Left Rotation Setting
: the rotation menu #2 will appear.
7.
Press button for
Left Angle
: adjust to display sequentially 0, 10, 15, 20, 25, 30, 35, 40 or max.
Left angle is the target left rotation angle to which the patient will turn to their left during rotation and
during left hold.
The rotation angle can be adjusted to only 0, 10, 15 or 20 if positioning packs are not used to
position patient during rotation. Refer to the
Lock-Out Menus
section in this chapter for additional
information.
If a left angle of 0
°
is selected, the right angle cannot be adjusted to 0
°
.
8.
Press button for
Left Pause
: adjust to highlight 0, 2, 5, 10, 20 or 30 min.
Left pause is the amount of time the patient is held in place once the target left rotation angle is reached.
9.
Press
Change
button to highlight either On or Off.
Current status indicates if rotation is activated or deactivated.
10.
Press button for
Enter
to save rotation settings. The
Acclimation Option
display shown at right will
appear.
Use acclimation option to help the patient adjust
to the target rotation angle by increasing the degree of rotation in a series of steps until the target
rotation angle is achieved.
For target rotation angles of 25
°
or more, the patient will rotate to 25
°
angle for six cycles, then
increase 10
°
angle every six cycles until the target rotation angles are reached.
If rotation is interrupted, the acclimation cycle will be restored at the last completed cycle if the
acclimation option has not been cancelled. Acclimation option is automatically cancelled when air is
turned off, or if CPR is activated.
Acclimation Mode will appear in the
Home
menu under
Percus: On Off
. This message remains on
the display after the completion of the cycles.
• Press
Yes
button to activate the acclimation option and return to
Home
menu.
• Press
No
button to return to
Main
menu.
ACCLIMATION OPTION
Do You Want to Approach the Programmed Angle in
a Series of Steps?
YES>
NO>