5-26
View Fowler and Trendelenburg Angles
1.
With the patient weight showing in the
Scale
display,
press
Scale On / Off
button shown at right.
The
Exit Scale
display will appear temporarily, as
shown at right.
2.
The
Fowler Angle Xº
display will appear, as shown
at right.
3.
To view Trendelenburg angle press either
Trend
button on the bed
motion control panel as shown at right.
4.
The
Trend Angle X
display will appear as shown at
right.
5.
Press
Scale On / Off
button to exit
Fowler / Trend
display.
Change Equipment
Use
Change Equip
button to hold current weight value in memory while other weight (traction, etc.) is
added or removed. Added / removed weight will not be reflected in weight reading.
1.
With the patient weight showing in the
Scale
display, press
and hold
Change Equip
button shown at right.
EXIT SCALE
FOWLER ANGLE Xº
Trend.
TREND ANGLE X
CHANGE
EQUIP.