Enterprise E8000X Service Manual
Servicing Instructions
Page 45
11.2.3 Connect the four actuator cables (“Electrical Assembly” on page 22, Item 15/16/17/18) and
OpenBus cable (“Electrical Assembly” on page 22, Item 19 and 20) to the control box, making
sure each cable is connected to the correct socket as previously noted. Thread the retainer
(Fig 10, Item A) over the cables and clip it into place.
11.2.4 Reconnect the backup battery (Fig 6, Item 4) to the control box, refer to Subsection 9.2.
11.2.5 Reconnect the mains lead (“Electrical Assembly” on page 22, Item 9) to the control box, refer
to Subsection 10.2.
11.2.6 Using the electrical safety test equipment (see “Tools and Equipment” on page 21), verify that
the earth bonding resistance value is within the limits shown in “Technical Data” on page 91.