Enterprise E8000X Service Manual
Preventive Maintenance
Page 7
CHAPTER 3
PREVENTIVE MAINTENANCE
This product has been designed to be virtually maintenance-free between service intervals. The degree
of periodic maintenance required will be determined by use and condition.
The following preventive maintenance checks and procedures should be carried out every 12 months.
All service and repair activities must only be carried out by properly qualified and trained persons
approved by Arjo.
Disposal of the product and its components should be compliant with local regulations.
1
General
Examine the bed for obvious signs of damage. Make sure that all visible fixings/fasteners are secure and
not missing.
Examine all flexible cables for cuts, abrasions, kinks or other deterioration.
Check the mains plug and cable for damage. If either the mains plug or cable is damaged, both must be
replaced as a complete assembly. Under no circumstances should the integral moulded plug be
replaced with a rewireable plug.
2
Castors and Brakes
Check the brakes for efficient operation.
Check that the steering lock operates effectively.
Poor braking or steering performance indicates that one or more castors require replacement (the
castors on this product are not adjustable).
Verify the condition of and tighten the screw S1306 - torque 10.2 Nm. See figure 5 Brake and Castor
Module in chapter 6 Servicing Instructions.
WARNING
Before starting any maintenance activity, disconnect the bed from the mains
power supply. The bed sections will still operate on battery power unless the
function has been disabled on the ACP.
Avoid skin contact with lubricants. Gloves and protective clothing should be
worn when carrying out maintenance work.
Summary of Contents for Enterprise E8000X
Page 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Page 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Page 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Page 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Page 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Page 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Page 104: ...This Page Intentionally Left Blank...