Enterprise E8000X Service Manual
Technical Data
Page 99
Recommended separation distances between portable and mobile RF
communications equipment and the bed
The bed is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the bed can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the bed as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
80MHz to 800MHz
800MHz to
2.5GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the power of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
d
P
1 2
.
d
P
1 2
.
d
P
2.3
Summary of Contents for Enterprise E8000X
Page 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Page 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Page 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Page 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Page 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Page 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Page 104: ...This Page Intentionally Left Blank...