Servicing Instructions
Enterprise E8000X Service Manual
Page 72
25
Roller Buffer - Replacement
This section describes how to replace the roller buffer. Reference is made throughout this section to
Figure 23 on page 68.
25.1
Removal
To remove the roller buffer:
25.1.1 Lift up the roller buffer moulding (Fig 23, Item 3). Undo the M5 Nyloc nut (Fig 23, Item 6) and
withdraw the M5 hex head bolt (Fig 23, Item 5) from the mounting tube.
25.1.2 Collect the shakeproof washer (Fig 23, Item 4) and remove the roller buffer moulding.
25.1.3 If necessary, pull the corner tube liner (Fig 23, Item 2) up and out of the tube.
25.2
Installation
To install the roller buffer:
25.2.1 If previously removed, fit the corner tube liner (Fig 23, Item 2) down into the mounting tube on
the bed extension fabrication (Fig 23, Item 1). Make sure the transverse holes in the tube and
liner are aligned.
25.2.2 Fit the roller buffer moulding (Fig 23, Item 3) over the tube.
25.2.3 Fit the shakeproof washer (Fig 23, Item 4) onto the M5 hex head bolt (Fig 23, Item 5). Fit the
bolt through the holes in the tube and liner.
25.2.4 Secure the assembly with a new M5 Nyloc nut (Fig 23, Item 6).
This procedure describes the replacement of the foot end roller buffer and/or
corner tube liner; use the same method to replace those at the head end.
Summary of Contents for Enterprise E8000X
Page 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Page 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Page 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Page 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Page 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Page 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Page 104: ...This Page Intentionally Left Blank...