9.1.3 Remove two panhead Pozidriv screws (Fig 6, Item 3) to detach the backup battery
(Fig 6, Item 4) from the mounting plate (E). Collect four Nyloc nuts (Fig 6, Item 1) and remove
the battery from the bed.
9.2
Installation
To install the backup battery:
9.2.1 Attach the battery (Fig 6, Item 4) to the mounting plate using two panhead Pozidriv screws
(Fig 6, Item 3) and four new Nyloc nuts (Fig 6, Item 1).
9.2.2 Refer to Figure 8. Connect the battery cable (C) to the control box (Fig 6, Item 5), making sure
the plug is pushed fully into the socket. Push down the retainer clip (B) to hold the battery cable
in place. Make sure that the battery cable cannot be accidentally pulled out of the control box.
WARNING
Make sure the battery test procedure is followed after installation refer to
Chapter 3, Subsection 3.