PiCUS TMSO Manual
69
9.2.3 Exchange of the TMSO. CU
I.
If you want to replace the TMSO.CU of an
active system with another one, e.g.
because the battery is too discharged and
the system has switched to energy saving
mode, you have to deactivate the system
first.
To do this, proceed according to point
and
System. All devices
are now in standby mode again.
II.
Since you want to reactivate the system
immediately with the new central and the
same project, you must now import the
current project and configuration data from
the old central into the TMSO.App so that
you can load it directly into the new central).
Switch to the
GET
tab in the Management
Window and click on the
GET
button under
Measurement Parameters
. The transfer
process is commented by corresponding
status messages.
CAUTION!
The read data are automatically transferred
to the project in the app and the
configuration of the central unit and
completely overwrite any existing data!
III.
You can then switch off the system by
proceeding as described in point