OP
ERA
TION
4.14
REM I/P OFF
EVENT
The external Remote Control input has opened
PRG COMPLET
EVENT
Program has completed
STOPPED
EVENT
STOP button has been pressed during a program
WARN LATCH
WARNING
A Sampler Warning has occurred and Special Option LATCH
WARNING OP is set for AUTO OFF or MANUAL
WARN T/OUT
WARNING
A latched Sampler Warning has automatically timed out
WARN DISAB
WARNING
A latched Sampler Warning has been manually disabled
TIMED OUT
WARNING
Sampler has failed to intake a sufficient volume of liquid into its Sample
Chamber within the allotted time (default 64 seconds)
See MAINTENANCE Section for details of how to react to this issue
CLEAN CHTOP
WARNING
There is a con
ductive path between the ‘common’
central Level
Sensing Probe and either one or both of the other two Level Sensing
Probes prior to commencing the sample intake phase
See MAINTENANCE Section for details of how to react to this issue
GUARD HIT
WARNING
The short Level Sensing Guard Probe has been contacted by incoming
wastewater before the opposite Normal Probe
See MAINTENANCE Section for details of how to react to this issue
OPEN CCT
WARNING
One or more of the 3 Level Sensing Probes are not connecting to its
contact springs at commencement of the sample cycle
See MAINTENANCE Section for details of how to react to this issue
BOT FAULT
WARNING
Electrical signals from Bottler are incorrect
See MAINTENANCE Section for details of how to react to this issue
T/SENS FLT
WARNING
The Sample Temperature Probe has failed or is not connected
B BCK-UP ON
WARNING
Appears where main power is lost with a Sampler being operated on
mains with a float charged Back-Up battery option
PWR RESTORD
WARNING
Appears where main power is restored with a Sampler being operated
on mains with a float charged Back-Up battery option
LOW VOLTS
WARNING
Battery has become fully discharged, or Mains Power supply has
dropped below acceptable limits
See MAINTENANCE Section for details of how to react to this issue
TRACT BLKD
WARNING
Some part of the Sample Tract is blocked causing an over pressure in
the Sample Chamber (where Blockage Detection System option is fitted)
MAINTENANCE Section for details of how to react to this issue
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......