5.6
Changing Desiccant Bag
It is important to ensure that electronic / electrical hardware inside the sealed Sampler
Module remains dry. As such periodically it may be necessary to change the Desiccant
Bag inside your Aquacell Sampling Module. To do this proceed as follows:
Switch off and isolate the Sampler module
Remove Sample Chamber / Top assembly (see removing / refitting the Sample
Chamber)
Remove rubber Desiccant Bag access plug from Sampler Module Front Panel
Withdraw Desiccant Bag and discard in suitable waste disposal bin
Remove new Desiccant Bag from its sealed bag and push into Front Panel
Replace the rubber Desiccant Bag access plug
Lubricating Pinch Valve Piston
Occasionally it may be necessary to lubricate the Pinch Valve Piston, this should be
carried out using the appropriate lubricant as follow *:
For ‘Pinch Valve Piston – Series 2’ (black) use ‘Silicone Grease – 25g’ (CL-6179) *
For ‘Pinch Valve Piston – Series 1’ (red) use ‘Pinch Valve Grease – Series 1 – 25g’
(CL-6140) *
* Caution – Using the incorrect grease is likely to cause damage to the Sampler Module
A small deposit of grease should be applied to the Pinch Valve Piston shaft and spread
over the working surfaces by manually moving it in and out
It is also a good policy to maintain the Sampler and its accessories in a generally clean
condition bearing in mind the potential infection hazards associated with wastewater
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......