OP
ERA
TION
4.16
5.
Set the Computer port properties as follows (settings may vary dependant on
communications program):
Bits per second
19,200
Data bits
8
Parity
None
Stop bits
1
Flow control
None
6.
Set the Computer ASCII Setup as follows:
Append line feeds to incoming line ends
Wrap lines that exceed terminal width
Emulation set to ‘Auto detect’
7.
Communication with the Sampler should now be possible
The Sampler can receive 7 control codes from the Computer, these are:
CODE
COMMENTS
s
[tagname]
Where [
tagname
] is a user definable tag / ID (up to twenty ASCII characters
long)
Example:
sSAMPLER_123 then carriage return gives the Sampler the
name SAMPLER_123
i
Shows the full list of codes, the computer will show:
and the Sampler LCD will show:
[1] TO DOWNLOAD
[2] TO ABANDON
i
Unit = SAMPLER_123
Date: 21-05-19 Time 15:30
Sampler firmware: SXX.XX.XX.XX
Temperature Correction = 0.0 C
Command list:
*i - Download Information
*e - Download Event Log
*w - Download Warning Log
*t1 - Download Temperature Log in Spread Sheet Format
*t2 - Download Temperature Log in Visual Appraisal Format
*s - Change unit name
*c - Cancel download
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......