OP
ERA
TION
4.7
SYSSAM5
SAMPLING MODE
TIME
[FLOW]
Samples can either be taken at fixed time intervals or,
when coupled to a suitable flowmeter, in relation to flow
➔
See SYSSAM8 if TIME
➔
See SYSSAM6 if FLOW
SYSSAM6
FLOW ORIDE TIME
DISABLED
[ENABLED]
When working with a FLOW signal, a forced timed sample
can be scheduled when insufficient effluent has passed to
activate a sample on flow basis alone
➔
See SYSSAM6
SYSSAM7
FLOW SIGNAL
4/20mA
[IMPULSE]
Sampler can accept either an impulse signal repeating at
equal flow increments or an analogue 4/20mA signal
corresponding to flow rate
➔
See SYSSAM8
SYSSAM8
REMOTE CONTROL
NO
[YES]
The sampling routine can be inhibited / enabled by an
external signal. Each time a new REMOTE CONTROL
signal is ON the sample routine will start, each time
REMOTE CONTROL is OFF the sample routine will stop
➔
See SYSSAM9 if YES and sampling by TIME
➔
See SYSSAM10 if YES and sampling by FLOW
➔
See SYSSAM11 if NO
SYSSAM9
CONTROL MODE
SAMPLER ENABLE
[PUMP/TIME TRIG]
With SAMPLER ENABLE selected, the start time for timed
samples will begin from zero each time REMOTE
CONTROL signal is switched to ON. NOTE: INITIAL
SAMPLE is defaulted to YES in this mode such that the
Sampler will always take an initial sample each time the
REMOTE CONTROL signal is switched ON
➔
See SYSSAM10
With PUMP/TIME TRIG selected, the start time for timed
samples begins when RUN PROGRAM is pressed and
sample timing continually counts from this point
irrespective of whether REMOTE CONTROL is ON or
OFF. NOTE: INITIAL SAMPLE is defaulted to YES in this
mode such that the Sampler will always take an initial
sample each time the REMOTE CONTROL signal is
switched ON
➔
See SYSSAM10
SYSSAM10
INITIAL DELAY
00HR 02MIN
A delay can be programmed in the Samplers recognition
of the external signal. This delay is typically used when a
pump is providing the control signal to allow the pump to
establish a representative flow
➔
See SYSSAM11 if Bottler is not fitted
➔
See SYSSAM13 if Bottler is fitted
SYSSAM11
OVERFILL PROT
NO
[YES]
The number of sample shots can be limited, when using a
single composite sample container to avoid overfilling
➔
See SYSSAM12 if YES
➔
See SYSSAM13 if NO
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......