3.7
Deploying the Sampler
Positioning
Aquacell Wastewater Samplers are NOT suitable for use in areas classified as hazardous
zones
The Sampler should be stood on a substantial, flat, level surface. It should be located as
close as possible to the sampling point and ideally at a level such that the Sample
Chamber outlet pipe is always above top water level of the liquid from which the samples
are to be taken
When choosing the position for the Sampler, consideration should also be given to the
correct installation of the Intake H
ose (see ‘
Routing
the Intake Hose’ section below)
Routing the Intake Hose
The Aquacell P2 Intake Hose is 2 metres long but if the required length is greater than 2
metres then one or more 2 metre Intake Hose Extensions can be added as required
The following diagram should be considered when routing your Intake Hose
Additionally the following points should be considered when installing your Intake Hose:
•
Ensure that there are no leaks in Intake Hose joints. Leaks will cause air bubbles in
the water being drawn up the Intake Hose and, in turn, the Samplers lift velocity will
be degraded
OK
✓
✓
NOT OK
Water trap
NOT OK
Potential
solids intake
OK
NOT OK
Potential siphon
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......