M
AINTEN
AN
C
E
5.9
5.
Display shows:
FAULT
SHOT TIMED OUT
and a shot is not acquired
Notes:
This message will only appear
briefly if a program is not in progress
when the fault occurs
This fault will not cause the program
to abort
a.
Intake filter is out of source
effluent
b.
Channel is dry
c.
Blocked sample tract
d.
See Symptom 2
a.
Ensure Intake Filter is fully
immersed. See Installing
Intake Hose (See Operation
Section)
b.
Ensure effluent is present in
channel whilst Sampling is in
operation
c.
Clean Sample Tract (see
Cleaning Sample Tract /
Stainless Steel Electrodes
earlier in this Section)
d.
No further action
6.
Display shows:
FAULT
BOTTLER FITTED
and a program is aborted
Bottler is connected to the
Sampler but not specified in the
program settings
Disconnect Bottler or reprogram
Sampler
7.
Display shows:
FAULT
BOTTLER MISSING
and a program is aborted
Bottler is not connected to the
Sampler but is specified in the
program settings
Ensure Bottler is fully connected
to Sampler then recheck
If fault persists contact supplier
for service advice
8.
Display shows:
FAULT
BOTTLER FAILED
and a program is aborted
Sampler is not receiving correct
signals from Bottler
Ensure Bottler is fully connected
to Sampler, then recheck
If fault persists contact supplier
for service advice
9.
Display shows:
FAULT - SAMPLE
TEMP SENSOR
and program is aborted
a.
Temperature Sensor is
connected to the Sampler but
not specified in the program
settings
b.
Temperature Sensor is not
connected to the Sampler but
is specified in the program
settings
c.
Electronic fault
a.
Disconnect Temperature
Sensor or reprogram
Sampler
b.
Connect Temperature Sensor
or reprogram Sampler
c.
Contact supplier for service
advice
10.
Display shows:
FAULT SAMPLE
TRACT BLOCKED
and Sample is aborted
Some part of the Sample Tract
is blocked causing an over
pressure in the Sample
Chamber during a pre or post
purge
Check and remove any
blockages from:
Intake Hose
Sample Chamber Top
Top Clamp breather orifice
11.
Display shows:
BACK-UP BATTERY
IN OPERATION
Appears where main power is
lost with a Sampler being
operated on mains with a float
charged Back-Up battery option
Restore mains power supply
Summary of Contents for Aquacell P2 Series
Page 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ...CONTENTS OF PACKAGE 1 2...
Page 14: ...CONTENTS OF PACKAGE 1 8...
Page 16: ...INTRODUCTION 2 2...
Page 26: ...INSTALLATION 3 2...
Page 36: ...INSTALLATION 3 12...
Page 38: ...OPERATION 4 2...
Page 56: ...OPERATION 4 20...
Page 58: ...MAINTENANCE 5 2...
Page 70: ...TECHNICAL SPECIFICATION 6 2...
Page 77: ......