![Applent Instruments Anbai AT8612 User Manual Download Page 71](http://html.mh-extra.com/html/applent-instruments/anbai-at8612/anbai-at8612_user-manual_2954893071.webp)
71
71
SCPI Command
Query Response:
<
float
>
Example:
SEND > OCP:MIN?
RET > 0.001
7.15
OPP subsystem
Table 7-11 OPP subsystem tree
OPP
:FILE
{FILE0,FILE1,…FILE9}
:RANGE
{3A,30A}
:VV
{FLOAT}
:VD
{FLOAT}
:START
{FLOAT}
:STEP
{FLOAT}
:SD
{FLOAT}
:END
{FLOAT}
:OV
{FLOAT}
:MAX
{FLOAT}
:MIN
{FLOAT}
OPP subsystem is used to set up OPP measuring parameter.
7.15.1
OPP:FILE
OPP:FILE Setting up OPP FILE.
Command Syntax:
OPP:FILE
< INT >
Parameter:
< INT >
Example:
SEND > OPP
:FILE 1
Query Syntax:
OPP:FILE?
Query Response:
<
INT
>
Example:
SEND > OPP:FILE?
RET > file1
7.15.2
OPP:RANGE
OPP:RANGE command sets the OPP CUR RANGE
Command Syntax:
OPP:RANGE
< INT >
Parameter:
< INT >
Example:
SEND > OPP
:RANGE 1
Query Syntax:
OPP:RANGE?
Query Response:
<
INT
>
Example:
SEND > OPP:RANGE?
RET > 3A
7.15.3
OPP:VV
OPP:VV command sets the OPP Start Voltage
Command Syntax:
OPP:VV
< float >
Parameter:
< float >
Example:
SEND > OPP
:VV 10.00
Summary of Contents for Anbai AT8612