SCPI Command
Query Response:
< float >
Example:
SEND > seq:cout?
RET > 10
7.10.5
SEQ:SAVE
SEQ:SAVE command set the sequence list file saving.
Command Syntax:
SEQ:SAVE
Parameter:
NONE
Example:
SEND > seq:save
7.10.6
SEQ:ERASE
SEQ:ERASE command sets the sequence list file erasing
Command Syntax:
SEQ:ERASE
Parameter:
NONE
Example:
SEND > seq:ers
7.10.7
SEQ:SET
SEQ:SET command sets the data of sequence list file
Command Syntax:
SEQ:SET
<step>, <level>, <width>
Parameter:
<step>, <level>, <width>
step: integer, which step
level:
single-precision
floating-point
format,
numerical value, unit according to operating mode
(
A,
V, W,
Ω
)
width: single-precision floating-point format, pulse
width, unit is "s"
Example:
SEND> seq:set 0,1.0000,0.01
// Setting up the data of
the first step as 1(A,V,W,Ω)
, 0.01s
Query Syntax:
seq:set?
Parameter
:
<step>
Step: integer, which step
Query Response
:
<level>,<width>
level:
single-precision
floating-point
format,
numerical value
width: single-precision floating-point format, pulse
width
Example:
SEND > seq:set? 0
RET > 1.0000,0.01
Restriction:
Can only be used in normal test status.
The execution of this command only aims at the operating
mode of the system being used, example: if system's
current operating mode is CC, then amend or query CC
list data.