AT8612 User's Guide
Pulse Mode
Under Pulse Mode, turn on load, in external trigger mode, each time the
instrument receives a low level trigger signal from trigger end at the rear
panel (means TRIG pin and GND pin short circuit ), the load will switch
to B value, after keeping pulse width B, it will return to A value. In internal
trigger, instrument receives trigger signal from TRIG at the front panel. In
remote trigger, instrument receives trigger signal from the remote control
side.
Trigger
Mode
Under [TRN] mode, in external trigger mode, each time the instrument
receives a low level trigger signal from trigger end at the rear panel (means
TRIG pin and GND pin short circuit ), the load will switch between A and
B value. In internal trigger, instrument receives trigger signal from TRIG
at the front panel. In remote trigger, instrument receives trigger signal from
the remote control side.
Under Trigger Mode, pulse width A and B are invalid, pulse width is
controlled by trigger signal.
3.5
<BAT MEAS PAGE> Page
AT861x can use constant current (CC) mode to test battery capacity.
Users can ensure battery capacity and discharge time by programming discharge current
and off-voltage, cooperate RS232/USB interface programming, users can see battery
discharge curves.
Figure 3-9 <BAT MEAS PAGE> Page
I
t
B = 10A
A = 5A
TRIG
0
4ms
4ms
TRIG
I
t
B = 10A
A = 5A
TRIG
0
TRIG
TRIG
TRIG