![Applent Instruments Anbai AT8612 User Manual Download Page 59](http://html.mh-extra.com/html/applent-instruments/anbai-at8612/anbai-at8612_user-manual_2954893059.webp)
59
59
SCPI Command
Parameter:
< float >
Example:
SEND >
basic:voff 5.00
Query Syntax:
basic:voff?
Query Response:
< float >
Example:
SEND
>
basic:voff?
RET
>
5.00
7.8.12
BASIC:TIME
BASIC:TIME command sets the on-load time.
Command Syntax:
BASIC:TIME
< float >
Parameter:
< float >
Example:
SEND >
basic:time 1.000
Query Syntax:
basic:time?
Query Response:
< float >
Example:
SEND > basic:time
?
RET > 1.000
Restriction:
Can only be used in conventional testing status
7.8.13
BASIC:VALUE
BASIC:VALUE
command sets the operating parameter.
Command Syntax:
BASIC:VALUE
<para>, <level>
Parameter:
<para>, <level>
para: instrument's operating mode
level: instrument's load parameter
Example:
SEND>
basic:value cc,1.0000 //constant current,
current value 1.0000A
Query Syntax:
basic:value?
Query Response:
< float, float, float, float >
Group 1's data: current value
Group 2's data: voltage value
Group 3's data: power value
Group 4's data: resistance value
Example:
SEND > basic:value?
RET > 18.0000, 1.0000,20.0000,1.0000
7.9
TRAN subsystem
Table 7-5 TRAN subsystem tree
TRAN
:CURRENT(CUR)
:VALUEA
<level>,<width>
:VALUEB
<level>,<width>
:VOLTAGE(VOL)
:VALUEA
<level>,<width>
:VALUEB
<level>,<width>
:POWER(POW)
:VALUEA
<level>,<width>
Summary of Contents for Anbai AT8612