AT8612 User's Guide
3.7
<SEQ VALUE SET PAGE> Page
Press [Setup] key, then press soft-key [SEQ], choose [FILE SET] to enter <SEQ VALUE SET PAGE> Page.
Figure 3-11 <Sequence List Effective Value Setup> Page
3.7.1
STEP 1 Setup
Steps of setting up Numerical Value of STEP 1
Step 1
Press[Setup] to enter <SETUP> page
Step 2
Press soft-key [SEQ ] to enter <SEQ MEAS PAGE> page
Step 3
Choose [FILE SET] field by using cursor key;
Soft-keys
Function
YES
Enter <SEQ VALUE SET PAGE> Page
Step 4
Choose [VALUE] field by using cursor key;
Step 5
Use numeric keypad to input value, press [Enter] to confirm.
Steps of setting up Timing of STEP 1
Step 1
Press[Setup] to enter <SETUP> page
Step 2
Press soft-key [SEQ ] to enter <SEQ MEAS PAGE> page
Step 3
Choose [FILE SET] field by using cursor key;
Soft-keys
Function
YES
Enter <SEQ VALUE SET PAGE> Page
Step 4
Choose [TIMING] field by using cursor key;
Step 5
Use numeric keypad to input timing value, press [Enter] to confirm.
* The setup of other STEP is the same as the above.
Example will be given to illustrate how to edit files:
Suppose:
Sequence list parameter: CC
Test Mode:
Continuous Mode
STEPS:
5 steps
1.
Step 1: 1.000A, 0.01s
2.
Step 2: 2.000A, 0.02s
3.
Step 3: 3.000A, 0.03s
4.
Step 4: 4.000A, 0.04s
5.
Step 5: 5.000A, 0.05s
6.
Save data in the File 9.