![Applent Instruments Anbai AT8612 User Manual Download Page 56](http://html.mh-extra.com/html/applent-instruments/anbai-at8612/anbai-at8612_user-manual_2954893056.webp)
56
AT8612 User's Guide
7.7.5
FETCH:RESISTANCE
FETCH:RESISTANCE command is used to retrieve the actual resistance value currently,
unit
Ω
Query Syntax
:
FETCH:RESISTANCE
Query Response
:
<
float
>
Example
:
SEND
>
fetch:res
RET
>
12.000
7.8
BASIC subsystem
Table 7-4 BASIC subsystem tree
BASIC
:MODE
{CC,CV,CP,CR}
:FUNC
{NRM,SHT,TRN,BAT,SEQ,ATF}
:RATE
{SLOW,MED,FAST}
:VMAX
{FLOAT}
:IMAX
{FLOAT}
:PMAX
{FLOAT}
:STATE
{ON,OFF}
:TRIG
{INT,EXT,BUS}
:4W
{ON,OFF}
:VON
{FLOAT}
:VOFF
{FLOAT}
:TIME
{FLOAT}
:VALUE
<PARA>,<LEVEL>
BASIC subsystem is used to set up the basic parameter of the instrument.
7.8.1
BASIC:MODE
BASIC:MODE command sets the instrument's operating mode.
Command Syntax:
BASIC:MODE <cc,cv,cp,cr
>
Parameter:
<cc,cv,cp,cr >
Example:
SEND > basic:mode cc
Query Syntax:
basic:mode?
Query Response:
<cc,cv,cp,cr>
Example:
SEND > basic:mode?
RET > cc
Restriction:
Can only be used in conventional testing status
Other command followed by this command will be ignored.
Example: basic:mode cc; *idn? // *idn? command will be ignored.
7.8.2
BASIC:FUNC
BASIC:FUNC command sets the operating function.
Command Syntax:
BASIC:FUNC <nrm,sht,trn,bat,seq,atf>
Parameter:
<nrm,sht,trn,bat,seq,atf>
Example:
SEND > basic:func nrm
Summary of Contents for Anbai AT8612