MT9090A/MU909011A Network Master User’s Guide
7-11
Measurement Accuracy of the Optical Power
Meter (MU909011A2 / A3)
Check that the measurement accuracy of OPM (Optical
Power Meter) satisfies the specification.
Note
Note
This test is only performed on units equipped with the
optional Optical Power Meter (OPM).
Connect the devices as shown in the following figure.
Figure 7-7: Optical Power Meter Measurement Accuracy Test –
Connection Diagram
Test Procedure:
Measure the measurement accuracy for input level -20 dBm
and measurement wavelength of 1550 nm.
Optical Fiber
Optical Fiber
MU909011
Series
Reference
Power Meter
Reference
Optical
Detector
Optical Variable
Attenuator
Reference Light
Source