MT9090A/MU909011A Network Master User’s Guide
7-9
Optical Power Level and Wavelength of Visual
Fault Locator (MU909011A1 / A3)
Check that the output power level and wavelength of the
optional Visual Fault Locator (the VFL is a Visible Laser
Diode) satisfy their respective specifications.
Note
Note
This test is performed only on units equipped with the
optional Visual Fault Locator (VFL).
Connect the devices as shown in the following figure.
Figure 7-6: Optical Power Level and Wavelength of VFL Test –
Connection Diagram
MU909011
Series
Optical Fiber (< 2 m, Bending radius
≥
50 mm)
Optical Spectrum Analyzer
Optical Sensor
Optical Power Meter