Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-119
TRG
Description:
Triggers a single sweep.
When single sweep triggering is selected for frequency, power level, or list sweeps,
a single sweep starts when the TRG command is received.
Functional
Category:
SWEEP
Related
Commands:
EXT
,
TRS
,
RSS
Front Panel Key:
Frequency
| Step Sweep | More > | Trigger > | Single | Trigger
Frequency
| List | Sweep > | Trigger > | Single | Trigger
Level
| Level Sweep | Trigger > | Single | Trigger
TRS
Description:
Triggers a single sweep.
Same as TRG. When single sweep triggering is selected for frequency, power level, or list
sweeps, a single sweep starts when the TRS command is received.
Functional
Category:
SWEEP
Related
Commands:
EXT
,
TRG
,
RSS
Front Panel Key:
Frequency
| Step Sweep | More > | Trigger > | Single | Trigger
Frequency
| List | Sweep > | Trigger > | Single | Trigger
Level
| Level Sweep | Trigger > | Single | Trigger
TSS
Description:
Advances sweep to the next step in dual step sweep mode.
The dual step sweep mode provides for generating synchronized, step sweep outputs for
two MG369xCs at a frequency offset. When operating in the dual step sweep mode, the
command, TSS, steps the sweep to the next point.
Functional
Category:
SWEEP
Related
Commands:
DU1
Front Panel Key:
N/A
TST
Description:
Starts an instrument self-test.
When TST is received, the MG369xC performs a self test then places a P (for pass) or a
F (for fail) on the bus. It also generates six self test results bytes. Extended Status Byte
one bit zero is set if self test fails; bit two is set when self test is complete.
Functional
Category:
SELF TEST
Output Command:
OSR
returns the six self test results bytes to the controller.
Front Panel Key:
System
| Self-Test
Summary of Contents for MG369 C Series