3-4 Commands
Command Directory
3-112
PN: 10370-10374 Rev. F
MG369xC GPIB PM
SNS <value> <unit>
Description:
Opens the step sweep number of steps parameter.
Permits setting the number of steps for the step sweep.
Functional
Category:
PARAMETER ENTRY
Value:
1 to 10,000
Units:
SPS
Related
Commands:
LIS
,
LGS
,
SDT
,
SSP
,
SWT
Output Command:
OSS
returns the number of steps of the step sweep to the controller.
Front Panel Key:
Frequency
| Step Sweep | More > | Number of Steps
SOF <value> <unit>
Description:
Opens the frequency offset parameter of the Secondary unit.
Permits setting the frequency offset parameter for the Secondary unit. The frequency
setting is determined by the frequency range of the particular MG369xC model.
Functional
Category:
PARAMETER ENTRY
Value:
Frequency range of the MG369xC model
Units:
GH, MH, KH, HZ
Front Panel Key:
Frequency
| CW | Primary Secondary > | Secondary Freqs. . . | VNA > | VNA Offset
SP0
Description:
Deselects non-equally spaced step sweep.
Functional
Category:
SWEEP
Related
Commands:
SP1
Front Panel Key:
N/A
Note
Use the SOF parameter entry command only to set the frequency offset for a Secondary unit that is
(1) Secondary to a Primary unit that is connected to a 360B VNA in a source or dual-source
configuration or (2) Secondary to a Primary unit that is programmed to perform non-sequential
frequency step sweeps.