Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-51
L5 or L5 <value> <unit>
Description:
Sets RF output power level to L5. Opens L5 parameter.
Causes the RF output power level to be set to the preset (or previously set) L5 power
level. Also permits setting the L5 parameter. The power level setting is determined by
the power level range of the particular MG369xC model.
Functional
Category:
POWER LEVELING
Value:
Power level range of the MG369xC model
Units:
DM (log), VT (linear)
Output Command:
OL5
returns the L5 power level (in dBm when in log mode; in mV when in linear mode)
to the controller.
Related
Commands:
LOG
,
LIN
,
PU0
,
PU1
,
PU2
Front Panel Key:
Level
|
Level
|
Level Control >
|
Level List. . .
| highlight
L5
, then press
Edit Selected
or
Output Level
L6 or L6 <value> <unit>
Description:
Sets RF output power level to L6. Opens L6 parameter.
Causes the RF output power level to be set to the preset (or previously set) L6 power
level. Also permits setting the L6 parameter. The power level setting is determined by
the power level range of the particular MG369xC model.
Functional
Category:
POWER LEVELING
Value:
Power level range of the MG369xC model
Units:
DM (log), VT (linear)
Related
Commands:
LOG
,
LIN
,
PU0
,
PU1
,
PU2
Output Command:
OL6
returns the L6 power level (in dBm when in log mode; in mV when in linear mode)
to the controller.
Front Panel Key:
Level
|
Level
|
Level Control >
|
Level List. . .
| highlight
L6
, then press
Edit Selected
or
Output Level