Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-91
PLW <value> <data>
Description:
Prepares the signal generator to receive a FM waveform.
A 2-byte value indicating the size of the waveform (minus 1) and 2-byte values containing
the waveform data must immediately follow the command. Each waveform value is a
16-bit, unsigned integer consisting of two bytes, the upper half of the value followed by
the lower half of the value. The size of the waveform must be a power of 2.
Serial command: 0x04ad
Functional
Category:
MODULATION
Example:
PLW<2-byte binary data for waveform size><2-byte binary data for
waveform data>
Related
Commands:
PHO, PMR0, PMR1, WFS
Front Panel Key:
NA
PM0
Description:
Turns off the Power Meter measurement mode
If the internal power meter option is not installed, this command produces a syntax error.
Functional
Category:
MEASURE FUNCTION
Related
Commands:
OPM
,
PM1
Front Panel Key:
System
| Meas. Power
PM1
Description:
Turns on the Power Meter measurement mode
RF power can be measured at an external source with a 560-7, 5400-71, or 4600-71 series
detector connected to the rear panel
POWER METER
connector. The measurement
results are sent to the controller upon receipt of the OPM command. Power
measurements can be taken simultaneously with modulation measurements. If the
internal power meter option is not installed, this command produces a syntax error.
Functional
Category:
MEASURE FUNCTION
Related
Commands:
OPM
,
PM0
Output Command:
OPM
Front Panel Key:
System
| Meas. Power