3-40 | IPS-4 Dual Bench (UV / IR) Analyzer
Adjusting the Zero/Span Fluid Flow Rate
When the zero or span (if used) flow solenoid is in an energized state,
Instrument Air (or Nitrogen or distilled water) flows into the sample
path via a solenoid manifold block and a flow meter. The flow rate of
Zero/Span (if used) fluid must be high enough to adequately flush the
Measuring Cell of sample fluid to obtain a “good zero.”
Typically, the IPS-4 Analyzer controls the zero (and span, if used)
flow rate; no adjustments are required.
If the customer provides the sample system zero, run the manual zero
at a flow rate of 2.5 L/minute (0.08 scfm). Since excess air flow into
the analyzer sample system will cool the Measuring Cell and Oven/
Sample System Enclosure, the optimum Zero fluid flow rate is the
lowest consistent with a good zero.
A good zero can be defined as a state where the concentration outputs
are stable and further increases in Zero fluid flow rate do not reduce the
concentrations observed on the analyzer User Interface Panel. A mid-scale
flow indicator (or meter) should be sufficient for this purpose.
While the sample system is warming up, or when the analyzer is in
standby mode (
Not Ready
state), zero flow may be desired. During these
conditions, it is important that the sample system is kept free of contami-
nation by the sample fluid.
For additional information, refer to the Sample System
Manual Supplement
in the “Supplemental Information” section of this manual or in the
Documentation Package shipped with the analyzer.
NOTE
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...