5-22 | IPS-4 Dual Bench (UV / IR) Analyzer
Purged Analyzers:
If it is necessary to work with the Electronics Enclosure door open
after replacing the Xenon Flash Lamp Assembly, take appropriate
precautions to avoid electrical shock. Hazardous voltages are present
inside.
Insert the key into the Purge Bypass Switch and switch it to the
“
BYPASS
” position (follow company policy). Follow any “Special
Conditions for Safe Use” listed in accompanying certifications.
Hazardous Location Applications:
Close the power-disconnect switch (breaker) to apply main AC Power
to the analyzer.
If heated Sample and Vent Lines are used, apply power to them.
11. Open the Vent Line Shut Off valve (in the Oven/Sample System En-
closure). Open the Sample Line Shut Off valve to allow sample fluid
into the sample system.
12. Close and secure the Electronics Enclosure doors, and allow the ana-
lyzer to warm up to operating temperature and stabilize. This could
take up to 2 hours if the sample system has cooled completely.
13. Zero and Span (if applicable) the analyzer.
14. Change the analyzer back to
Measure
mode (from the
Calibration
screen, select
Start measure
and press
ENTER
). Press the
HOME
soft
key to return to the
HOME
screen.
15. Adjust the Aspirator Pressure Regulator to achieve normal operating
flow rates, as indicated on system Final “As-Built” drawings. The Flow
Indicator (if used) should be in the centre of the meter for an optimal
flow rate.
The procedure is complete.
!
WARNING
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...