Specifications
| 2-9
DNV GL Presafe AS,
Veritasveien 3, 1363 Høvik, Norway
2 of 3
EU-Type Examination Certificate
[13]
Schedule
[14] EU-TYPE EXAMINATION CERTIFICATE No.:
Presafe 15 ATEX 7029X
Issue 1
[15] Description of Product
The IPS-4-DB is a dual-bench process spectrophotometer that is housed in a two part IP 65 rated stainless
steel enclosure, and has a certified purge control system, an Expo MiniPurge Model 07 1XLC/ss/PO or Model
07 1XLC/SS/ET/PO, SIRA 01 ATEX 1295X on the part of the enclosure that houses the electronics.
The sample system is housed in the unpurged part of the enclosure and contains an Ex d heater. The power
is supplied into the Ex px enclosure from a component approved Ex d enclosure that contains rel ays to cut all
of the circuits going into the purged enclosure via ATEX certified bushings. The relays for the purged portion
of the enclosure are controlled by an Ex d pressure switch that is operated by the purge controller. The
pressure relief valve on the pressurized enclosure is part of the certified purge control system and is set at 1
000 Pa. An optional Ex d certified pressure transmitter (Rosemount) is available for monitoring the sample
system pressure and is simply connected to the IPS-4 through an existing entry on the disconnect enclosure.
Type designation
IPS-4-DB
Electrical Data
120/240 VAC 50-60 Hz, 850 W
Degrees of protection (IP Code)
IP65
Ambient temperature:
-20°C to +50°C
Routine tests
Routine testing is required per EN 60079-2 clause 17.1 and 17.2.
Purge data
Minimum Purging Flow Rate..................... 225 l/min
Minimum Purge Time............................... 5 Minutes
Protective Gas............................................ Air or Nitrogen
Minimum Supply Pressure............................ 4.15 bar (60psi)
Maximum Supply Pressure........................... 5 bar (72.5psi)
Maximum Leakage Rate........................... < 15 l/min
Minimum Overpressure.............................. 50 Pa (0,5 mBar)
Maximum Overpressure............................. 1000 Pa
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...