Maintenance and Troubleshooting
| 5-15
Use an o-ring removal tool to remove the outer o-ring from the
Cell. Take care not to scratch or damage the o-ring groove.
Carefully remove the Cell Window from the Cell by grasping it by
its outer edge and lifting it out of the Cell. Avoid touching the win-
dow surfaces. Set the Window aside on a soft, non-abrasive cloth.
d. While holding the Cell vertically so that the Reflector Block end is
up, remove the three M4 x 20 mm screws that secure the Reflector
Block to the Cell.
Remove the Reflector Block carefully to ensure
the Cell Window behind it does not drop out.
e. With the Reflector Block end up, carefully remove the o-rings and
Cell Window from the Cell.
Inspect the Windows for any chips, cracks, or scratches. If they are
not damaged, they can be cleaned and reused; set them aside on a
soft, nonabrasive cloth.
If the Windows are suitable for reuse, use Kimwipes
®
EX-L or an
equivalent extra low-lint tissue to clean the front and back sides. A
high purity solvent, such as Isopropanol (99 %), may also be used
(if suitable for use with the sample species). Rinse with pharma-
ceutical-grade distilled water.
Replace chipped, cracked, or scratched Windows when reassembling
the Cell.
f.
Systems with a Cell RTD:
Note the position of the Cell RTD Clamp (see Cell RTD in Figure
5-1) and remove it from the Cell.
g. Inspect the interior of the Cell Body for foreign material. If neces-
sary, use a cotton swab to clean the Cell Body.
If a more thorough cleaning is required, use only cleaning materi-
als that are suitable for use with the sample species and the Cell.
Clean the fittings using the same method as the Cell Body.
For example:
If suitable for the species, an acceptable material is Chem-Thane –
or a nonabrasive detergent and water solution. The Cell Body can
then be rinsed with reagent-grade acetone (if suitable), followed
by a rinse with pharmaceutical-grade distilled water.
h. Allow all components to dry thoroughly before reassembling. Dry,
oil-free instrument air can be used to remove any remaining mois-
ture from the passages in the Cell Body and fittings.
!
CAUTION
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...