Installation and Start-Up
| 3-21
To make the signal, relay contact, and AC power connections:
1. Open the Disconnect Enclosure or Sample Line Disconnect Box.
If a power-disconnect switch (breaker) was installed, open (turn off)
the breaker before making the AC power connections. AC Power and
relay connections are made at the AC Disconnect board.
2. Route the analog input/output signal wiring and alarm relay contact
conductors through one of the Signal cable entry ports and into the
Disconnect Enclosure.
Terminate the signal and alarm relay contact wires as detailed in the
Interconnect Wiring drawing
(part of the analyzer Final “As-Built”
drawings package).
3. Route the AC power conductors through the Power cable entry port
and into the Disconnect Enclosure.
Terminate the AC power wires as detailed in the Interconnect Wiring
drawing.
Do not apply power to the system until after all of the wiring has been
installed, connected, and verified, and only if the purging system (if
used) is ready for operation.
4. Plug all unused cable entry ports with certified plugs.
Permanent Ethernet Cable Connections (Optional)
When using the Ethernet interface on the analyzer, the minimum re-
quirements for the Ethernet cable is a CAT 5e (maximum 100 m) cable to
connect the analyzer to the network, up to a maximum of 100 m from
the hub. For outdoor analyzer installations, hazardous locations, and/
or lengths greater than 100 m, special requirements may be necessary.
Contact your network administrator for information.
For details about setting up Ethernet parameters, refer to “Ethernet
Network Screen” in Chapter 4.
To install the customer Ethernet cable:
1. Route the Ethernet cable through one of the Signal cable entry ports
and into the Disconnect Enclosure.
2. Connect the cable to
J6
.
3. Route the cable to the network connection in the plant and connect it.
NOTE
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...