Installation and Start-Up
| 3-35
Start-Up
Diagnostic Checklist
This checklist describes problems that can be encountered while powering
up the analyzer and/or preparing it for normal operation. If the analyzer
experiences problems during power-up, review this checklist and perform
the corresponding corrective action to fix any problems.
Problem Encountered
Status/error code conditions for current alarms are displayed on the
User Interface Panel.
Corrective Action
To view current alarms, press the
ALARMS
soft key to view the
ALARMs
screen. For information about viewing current alarms, refer
to “Alarm Notification (HOME Screen)” and “ALARMS Menu Screen”
in Chapter 4. For detailed information about these alarms, refer to
the alarm conditions and corrective action in the “Maintenance and
Troubleshooting” chapter of this manual.
Problem Encountered
The
Xenon Flash Lamp
is not operating (intermittent clicking sound is
not heard).
Corrective Action
Check the system state on the right side of the HOME screen. Note
that the Xenon Flash Lamp will not operate when the system is in the
Not Ready
state.
If the system is in a state other than
Not Ready
, take appropriate
safety precautions and check for proper wiring connections.
Problem Encountered
The LEDs on the
Relay
,
Customer Connection
, or
MCU
boards are
not on.
Corrective Action
Take appropriate safety precautions and check for proper I/O wiring
connections.
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...