Maintenance and Troubleshooting
| 5-23
Replacing the Infrared (IR) Source Assembly
The typical life expectancy of the
Infrared Source
is 3–5 years of continu-
ous operation. However, if the
Infrared Source
begins to exhibit signs of
degradation prior to the 3-year mark, replace the
Infrared Source
Assem-
bly (Figure 5-3.2). See Figure 5-1 for the location of this assembly in the
Electronics Enclosure.
Refer to “Optical Bench Assembly Spare Parts” in Chapter 6 for reor-
dering part numbers.
NOTE
Figure 5-3.2.
Infrared (IR) Source
Assembly (100-2847).
To replace the Infrared Source Assembly:
Hazardous Locations
Before proceeding, test the area around the analyzer for flammable
gases. If an explosive gas atmosphere is present, do not power down
the analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
1. Manually Zero the analyzer for 10 minutes.
2. After the Zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
!
WARNING
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...