3-42 | IPS-4 Dual Bench (UV / IR) Analyzer
Setting Sample Response Time
Sample response time can also be used as a parameter for setting the
Aspirator Pressure
. Response time is determined by switching the analyz-
er from
Zero Gas Flush
mode to
Measure
mode and recording the time it
takes the analyzer to display the first reading after the switch.
To check the sample response time:
1. Observe the output concentration from the
HOME
screen.
2. Press the
SETUP
soft key to display the
SETUP
menu and then press
the
DOWN
arrow key until the
Calibration
sub-menu item is displayed
and selected.
3. Manually zero the analyzer. To do this, press
ENTER
to display the
Calibration
screen. Use the
DOWN
arrow key to select
Start zero
calibration
and press
ENTER
. The message “
Please wait ...
” will appear
while the analyzer initializes and performs the calibration function.
Run the manual zero for 10 minutes. For more information about set-
ting the Zero fluid flow rate, refer to “Adjusting the Zero/Span Fluid
Flow Rate” in this chapter.
4.
From the
Calibration
screen, select
Start measure
and press
ENTER
to
return the analyzer to Measure mode.
5. Press the
HOME
soft key to return to the
HOME
screen and observe
the analyte output concentration(s) again.
If the sample response time is adequate (first response after a Zero is
less than 30 seconds to T90), no further adjustment is necessary.
The sample response time may vary, depending on Sample Line
length.
NOTE
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...