Maintenance and Troubleshooting
| 5-21
5. Disconnect the
DC Power Cable
connector (from the Lamp Housing)
at
J2
on the
Xenon Lamp
board located near the top of the Electronics
Enclosure (see Figure 5-1 for location).
6. While holding the Lamp Assembly with one hand, loosen (by hand)
the blue knurled nut that secures it to the Optical Bench Assembly
and remove the Lamp Assembly.
Dispose of the
Xenon Flash Lamp
Assembly
following company policy, or local regulations. Refer to
“UV Source Lamps Disposal” in the contents section of this manual
for more information on disposal recommendations.
Do not use any tool to loosen the blue nut on the Optical Bench when
removing or replacing the Xenon Flash Lamp Assembly.
7. Take the new Xenon Flash Lamp Assembly (Part No. 100-2761) and,
using the alignment pin on the Optical Bench Assembly as a guide,
gently push the Lamp Assembly straight into place.
Tighten, by hand, the blue knurled nut to secure the Lamp Assembly
to the Optical Bench Assembly.
8. Connect the
DC Power Cable
from the Lamp Housing to
J2
on the
Xenon Lamp
board.
9.
GP Analyzers Only:
Turn on the Instrument Air supply to the analyzer and open the Aspi-
rator Air valve.
10. Take all necessary precautions to safely power up the analyzer and
prepare it for operation:
Do not apply main AC power to the analyzer if the Xenon Flash Lamp
Assembly electrical connections have not been made.
Before proceeding, test the area around the analyzer for flammable
gases. If an explosive gas atmosphere is present, do not apply power to
the analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
When the analyzer doors are open, take appropriate precautions to
avoid electrical shock. Hazardous voltages are present inside.
!
CAUTION
!
WARNING
!
WARNING
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...