Installation and Start-Up | 3-19
• Use the conduit entry point closest to the connections you are making.
Do not add any additional conduit entry holes!
Control Unit AC Mains Supply Connections
Follow general grounding and shielding requirements for all wiring
as described in the “General Wiring Requirements” section of this
Wiring section.
Do not run control unit AC mains supply wiring in the same conduit
with other AC mains supply wiring. By keeping this wiring separate,
you prevent transient signals from reaching the control unit.
The Series 2000 control unit can operate using between 95-to-230 volts AC,
47-to-63 Hertz. There is no power switch or circuit breaker on the control
unit. It must be protected by installing it on a circuit-protected line with
maximum 15 amperes, a switch or circuit breaker in close proximity to
the control unit, and within easy reach of an operator. Mark the switch or
circuit breaker as the control unit disconnecting device.
Power connections to the control unit are as follows:
Line
Line connection
Neutral
Neutral connection (USA)
Chassis Stud
Equipment ground (protective conductor)
Use the ½” conduit entry hole in the control unit for AC mains supply
wiring. Use the chassis stud next to the ½” conduit entry hole for equip-
ment ground (protective conductor).
SENSOR TYPE
# of TWISTED
PAIRS NEEDED
GAUGE WIRE
TM2000
4
22 (to 250 ft)
18 (250 to 1000 ft)
Figure 3-17.
Recommended sensor
cable wiring (minimum
acceptable grades).
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...