Calibration | 5-21
Set Cal Timers - Set Auto Timers
This menu option allows you to set the frequencies for automatic calibra-
tions or verifications, including the time when the first calibration or veri-
fication starts. You have the option of setting the cycle in either hours or
days. Note that the auto calibration cycle must also be enabled to perform
an automatic calibration or verify (Auto Cal On/Off).
If you schedule a verification and calibration at the same time, the
verification will be performed first, then the calibration.
If using a digital input to start an auto calibration or auto verify, you
must select calibrate or verify from the Remote Switch Configuration
option from the Setup key menu.
1. Select
Set Auto Timers
from the Set Cal Timers menu.
2. Select
Auto Cal Cycle
if defining an automatic calibration cycle. Se-
lect
Auto Verify Cycle
if defining an automatic verification cycle.
3. Select
Set Cycle in DAYS or Set Cycle in HOURS
, depending on
whether you want to set the frequency of calibrations or verifications
in days or hours.
If you are setting the auto timers for the first time, Series 2000 software
will automatically prompt you to enter the current time and date after
timers are set up.
You will then be prompted to enter the frequency for the calibration or
verification cycle (either in days or hours, depending on what you se-
lected in step 3 above):
Cal Cycle = 01 days
New Value = xx
4. Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
NOTE
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...