5-26 | Thermox TM2000 Oxygen Analyzer
You will be prompted to enter the verify gas value:
Verify Gas = xx O2
New Value?
3. Enter the verify gas value, then press the
Enter
key. If you wish to
keep the current verify gas value unchanged, press the
Enter
key
without entering a value.
4. Enter the acceptable verify range. For example, if you enter a verify
gas range of 1%, and the verify gas value is 10%, then the acceptable
range is between 11% and 9%.
If during a verify operation (Initiate Verify menu option) the system
finds that the verify gas is not within the acceptable range, a Verify
Failure message will appear on the bottom line of the display, and
the Service Alarm will trip. The error message will continue to be
displayed until a successful verify is completed (gas value within the
range) or a successful calibration is completed.
NOTE
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...