5-6 | Thermox TM2000 Oxygen Analyzer
See the “Remote Calibration Unit Connections” section in the Installation
chapter for help on how to trigger an automatic calibration using a digital
input. When using a digital input to trigger the auto calibration, you must
make sure that the Setup Key/Remote Switch Config selection is set to
calibrate, and not verify.
If you replaced a cell, or your cell has failed a regular calibration
because of a deteriorating cell, you will need to perform a primary
calibration rather than a regular calibration. See the “Primary Cali-
bration” section in the Chapter 4 for help.
See the Chapter 3, Installation, for calibration gas requirements.
Auto Calibrate
• Enter the calibration gas value to match the zero calibration gas cylin
-
der you intend to use to calibrate your system.
• Define how long you want the calibration gas to flow through the sen
-
sor (Set Cal Timers/Cal Gas Duration). Remember that with an auto
calibration, the RCU automatically shuts down the zero calibration
gas based on the calibration gas duration time you specify. Be sure to
allow enough time for the gas to stabilize. You may want to use the
“Inject Cal Gases” option from the Calibrate menu to determine how
long it takes for the gas from the remote calibration unit (RCU) to
reach the sensor and stabilize and then use this time.
• Define a time period for the system to recover from reading the cali
-
bration gas value to reading process gases (Set Cal Timers/Recovery
Duration).
Start the auto calibrate as follows:
1. Select the Initiate Cal option from the Calibrate key menu.
2. Select Auto Calibrate.
NOTE
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...