5-18 | Thermox TM2000 Oxygen Analyzer
Inject Cal Gas
This menu option requires that an auto cal card and RCU are in-
stalled.
This menu option allows you to manually adjust the pressure and flow
of your calibration gas (pressure of calibration gas cylinders and flow into
sensor) by cycling through the solenoid valves on the RCU. No calibration
or verification data is collected while in this menu option. See the “Me-
chanical Installation” section in the Installation chapter for recommended
calibration gas pressure and flow requirements.
It is important that all calibration gases, as well as the sample gas, flow
through the analyzer at the same flow rate.
1. Connect the span and zero gas cylinders to the RCU.
2. Select the
Inject Cal Gas
menu option from the Calibrate key menu.
The following message will then appear on your display, indicating that
the RCU has introduced the span gas to the sensor:
Span/Verify gas is flowing
You will be prompted to press the
Enter
key once you have manually set
the pressure and flow for the span gas:
ENTER for next Gas.
CANCEL to quit.
3. When you have adjusted the pressure and flow for the span gas to the
desired levels, press the
Enter
key.
NOTE
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...