5-4 | Thermox TM2000 Oxygen Analyzer
• Temporarily enable or disable auto calibrations or verifications (Set Cal
Timers/Auto Cal On/Off).
• Define how long calibration gases should be applied to the sensor dur
-
ing an auto calibration or verification (Set Cal Timers/Cal Gas Dura-
tion).
• Define recovery duration times (Set Cal Timers/Recovery Duration).
• Define verify gas values to match known values of verify gas cylin
-
ders; enter acceptable verify range (Configure Verify).
Aborting a Calibration
If you press the
Calibrate
key while the system is performing a calibration,
you will be have the option to either abort or to continue the calibration.
All other menu functions are temporarily disabled during a calibration
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...