Controller / User Interface | 4-25
Set Current Mode Option
The
Set Current Mode
menu option allows you to define either a 0-to-20
or 4-to-20 mA analog output range for the selected analog output port.
This output is dependent on both the range of readings you define (see
the “Set Current Range” section), and on the function you track on the
analog output.
Setting the current mode
1. Select the
Set Current Mode
menu option from the Analog Range Key
after selecting an analog output port.
2. Select
4-to-20 mA Mode
or
0-to-20 mA Mode
.
The following will then appear on the display:
Out # y is set for
4 - 20 ma
or
Out # y is set for
0 - 20 ma
Output Filtering Option
The Output Filtering menu option allows you to filter out quick transi-
tions in readings to allow a smoother analog output.
Setting output filtering
1. Select
Output Filtering
from the Analog Range Key after selecting an
analog output port.
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...