4-20 | Thermox TM2000 Oxygen Analyzer
(Select Function). Choices are O
2
readings, cell temperature, cell mil-
livolts, and thermocouple millivolts.
Once you select what type of process readings the analog output
should track using the Select Function menu option, decide what
range of these readings the analog output range will cover (Set Cur-
rent Range).
Decide whether to enable output filtering, and if enabled, how much
to smooth out abrupt changes in readings (Output Filtering).
Analog Range Key menu options are presented in the order in which
they appear on your control unit screen, from top to bottom. Figure 4-5
provides an overview of the functions covered by each Analog Range Key
menu option.
Figure 4-5.
Overview of Analog
Range Key functions.
ANALOG KEY
MENU OPTIONS
FUNCTIONS/OPTIONS
Set current Range
Range of readings that the analog output represents
(for example: 4% o
2
= 4 mA; 10% O
2
= 20 mA).
Set Track/Hold
Hold output at last process reading during a cal or
verify
Set current Mode
0 to 20 mA
4 to 20 mA
output Filtering
Smooth out variations in readings
Select Function
o
2
cell Temperature
cell Millivolts
Thermocouple Millivots
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...