Serial Communications | A-5
Development Serial Communication Commands
These commands are used during initial serial communication design.
Echo (A)
The echo command is used to test the serial protocol and line integrity.
Any text sent to the Series 2000 Controller is echoed back to the host.
Bad Command (B)
Just returns a bad command (error code 01) reply.
Acknowledge (C)
Returns a simple acknowledge. This command can also be used later to
check that the controller is on-line and ready to communicate.
Data Format (J)
This command is used to check how data is output through the serial link.
For each variable that is accessible through the Read Number and Write
Number commands, a string is returned that describes the format of the
number. This response data consists of three characters:
Format
Access
Location
Where
format
is:
Access
is as follows:
F
Float
r
read only
H
Hexadecimal
w
write only
U
Unsigned decimal
b
read/write
Location
is:
r
RAM
e
EEPROM
Due to the nature of EEPROM memory, each location in the memory has
a limited number of writes allowed (approximately 10,000). The location
character can be used to determine where the data is stored internally to
the Series 2000 Controller and act accordingly.
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...