Calibration | 5-25
Set Cal Timers - Recovery Duration
This menu option applies to all automatic, remote, and manual cali-
brations and verifications.
This menu option allows you to define a recovery time so the control unit
can return to reading process gases after reading calibration gases without
affecting the analog outputs or the triggering of alarms.
If, for example, your process gas is 6%, and your last calibration gas is
2%, this option gives the system time to recover back to reading the 6%
process gas. This menu option also can be used to prevent triggering of
alarms if, for example, your low alarm is set to activate if readings fall
below 3%, and your calibration gas is 2%.
1. Select
Recovery Duration
from the Set Cal Timers menu.
You will be prompted to enter a calibration recovery duration:
Cal Recovery = xx:yy
Enter new time: Mn:Sc
2. Enter the recovery time, then press the
Enter
key. If you wish to use
the currently displayed recovery time, press the
Enter
key without
entering any text.
Configure Verify
The Configure Verify menu option allows you to define the verify gas
value and the acceptable range of readings for a verify.
To set your verify gas value and acceptable verify range, do the following:
1. Select Configure Verify from the Calibrate Key menu.
2. Select whether you wish to enter the verify gas using a percent or
PPM scale.
NOTE
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...