3-4 | Thermox TM 2000 Oxygen Analyzer
Sample gas / calibration gas requirements
The maximum inlet temperature of the sample gas (or calibration gases)
is 70° C (160° F). The sample gas must also be clean and dry. If the sample
contains any condensing moisture, you should heat trace the entire sam-
ple line to prevent condensation. Any condensing moisture in the sample
can damage the zirconium oxide cell in the sensor!
Figure 3-2.
Sensor mounting
(with sample bypass).