6-6 | Thermox TM2000 Oxygen Analyzer
Temp Rise Failure
This indicates that the sensor has failed to increase a minimum of 10° C in
a 60-second time frame during start-up. Once the system reaches the op-
erating temperature set point, this message will appear only if the sensor
falls below the set point by 15° C and fails to recover within 60 seconds.
The problems that can cause this error are as follows:
AC line power problem at the sensor
Open furnace
Over-temperature protection circuit trip on the sensor board
Sensor board failure
Loss of furnace drive signal from control unit
Thermocouple failed (shorted or opened)
Interconnecting wiring problem
Polarity of thermocouple is incorrect
T/C (Thermocouple) Circuit Failure
This message appears when the control unit recognizes a sudden drop in
temperature (greater than 100° C). The thermocouple circuit failure mes-
sage can be the result of one of the following problems:
Shorted thermocouple
Faulty interconnecting wiring
Control unit display module failure
Sensor board failure
If this message is soon followed by a “Thermocouple Failure” message, see
that message for troubleshooting instructions.
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...