A-2 | Thermox TM 2000 Oxygen Analyzer
Command Message Format
The command message format is as follows:
Where:
The
Start Character
is the greater than character (“>”, which is ASCII
decimal character 62 or hexadecimal character 3E).
The
Node Address
is from 00-FF, and is represented as two hex ASCII
characters. Each controller has a unique node address (see the “Communi-
cation/Node Address” section in the Setup Key chapter for help on speci-
fying node addresses for controllers). Note, however, that the maximum
number of controllers that can be connected on the network is 32.
The
Command Letter
is from ASCII A-Z (upper case only).
Data is a string of characters whose meaning depends on the command
(see below) and where the number of characters returned also depends
on the command. Limit is 20 characters.
The
Checksum
is the sum of all the preceding characters (excluding the
>> character) - modulo 256.
The
End Character
is a carriage return (equivalent to ASCII decimal char-
acter number 13).
The system disables any checking of the checksum if the checksum
character is equal to ?? (?? is equivalent to 2 ASCII 63 characters).
This makes testing by the programmer easy, but it is strongly recom-
mended that the integrity check is enabled in the final system.
NOTE
Start Character
Node Address Command
Letter
Data
Checksum
End Character
“>” (AScii deci-
mal character 6)
2 hex Ascii
characters
(00-FF)
(See com-
mands sec-
tion for list of
commands
(n characters
- depends on
command)
2 hex AScii
characters
(sum except
the “>” char-
acter
cariage return
(AScii decimal
13)
Summary of Contents for Thermox TM2000
Page 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Page 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Page 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Page 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Page 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Page 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Page 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...