Temperature control
240
Goldeye G/CL Technical Manual V4.2.0
•
If the
TemperatureControlTarget
of the camera
is
activated, the camera
heats the sensor and levels the temperature at the setpoint of +35 degrees
Celsius.
Operational statuses
explains the operational statuses that are indicated by the temperature
status LED on the camera back panel.
For TECless cameras, the LED will either remain off, not indicating any TEC status;
or it will turn red, indicating overheating.
Figure 77: Heating capabilities of Goldeye G/CL-008 models (example)
Use an efficient heat removal device
If the camera has switched to
Alert
status, it is also recommended to provide a
more efficient heat sink for the camera housing before resuming operation.
However, additional heat dissipation arrangements are always advantageous.
LED color
flashing pattern
Status and description
Off
There are two cases when the LED remains off:
•
The TEC is actively switched off, using the feature
SensorTemperatureControlMode
.
•
The camera is a TECless camera that has no TEC
available.
Off
Deviated
The sensor cooler is operating, the defined sensor
temperature (setpoint) has not been reached or stabilized
yet. This signal is no error signal.
Green-red
(flashing)
Table 145: Temperature status LED (sheet 1 of 2)
-21 °C
-10 °C
0 °C
10 °C
20 °C
30 °C
40 °C
50 °C
Ambient temperature
Default setpoint (sensor)
Exis
Ɵ
ng setpoint (sensor)
Housing temperature
Desired setpoint not reached
TemperatureControlTarget
not ac
Ɵ
vated:
only low setpoint reached
TemperatureControlTarget
ac
Ɵ
vated:
desired setpoint reached