Filters and mounts
174
Goldeye G/CL Technical Manual V4.2.0
Removing the filter from the mount
1. On the back side of the mount, loosen the three cylinder bolts (5) of the filter
fixing ring (4), using a Torx T6 tool.
2. Remove the fixing ring.
3. Remove the existing optical filter (3) from the adapter.
Installing a filter into the mount
1. Insert the new optical filter into the matching recess.
-
See
(blue = filter, red = filter recess).
-
See
for dimensions of suitable optical filters.
2. Insert the fixing ring into the adapter.
3. Fix the ring with the three cylinder bolts.
-
Tighten the bolts with a maximum torque of 0.25 Nm.
Attaching the mount to the camera
1. Screw the adapter into the camera.
2. Tighten the locking screw with a maximum torque of 0.1 Nm.
Figure 48: Changing the filter in the M42-Mount adapter
2
3
4
5
1